AUTOPOL IV Polarimeter
Manufacturer: Rudolph Research Analytical
SKU: Inquire
Description
Rudolph Research Analytical's AUTOPOL IV is a microprocessor-controlled automatic polarimeter available in configurations offering one, two, or six measurement wavelengths. It measures optical rotation, specific rotation, and concentration, with a broad measuring range of +/-89° for optical rotation, +/-999.99° for specific rotation, and 0 to 99.9% for concentration. The instrument achieves an accuracy of 0.002° up to 1°, 0.2% up to 5°, and 0.01° above 5°. Users can select from several wavelengths, including 365 nm, 405 nm, 436 nm, 546 nm, 589 nm, and 633 nm (with 578nm and 880nm also available). The temperature range for sample measurement is from 10°C to 100°C. Features include a touchscreen interface, customer-defined programs for simplified operation, temperature display and correction, statistical functions, and data communication via RS-232 and parallel printer ports. Its versatility makes it suitable for various research and quality control applications in chemistry and pharmaceuticals.
Specifications
| Item | AUTOPOL IV Polarimeter |
| Company | Rudolph Research Analytical |
| Catalog Number | Inquire |
| Measuring Range | +/-89° (Arc Optical Rotation), +/-999.99° (Arc Specific Rotation) and 0 to 99.9% (Concentration) |
| Measurement Mode(s) | Optical Rotation, Specific Rotation, Specfic Rotation Plus Concentration, User Defined Scale, °Z Sugar Degrees |
| Accuracy | 0.002° up to 1°, 0.2% up to 5° and 0.01° above 5° |
| Measurement Wavelength | 365 nm, 405 nm, 436 nm, 546 nm, 589 nm, and 633 nm (or 578nm and 880nm) |
| Temperature Range | 10°C to 100°C |
| Quantity | EA |
| Type | Benchtop |
| Resolution | 0.001° Arc Optical Rotation (or 0.001% Concentration or 0.001 Specific Rotation) |
| Sample Volume | 200 mm |
| Measuring Time | 4° per sec slewing rate and 5 sec nominal settling time |
| Dimensions | (WxDxH) 30 x 17 x 11.5 in (or 762 x 279.4 x 431.8 mm) |
| Optical Wavelength | 365 nm, 405 nm, 436 nm, 546 nm, 589 nm, and 633 nm |
| Temperature Compensation | 15°C to 40°C (±0.2°C) |